ISO 22412:2025: changes in DLS and PdI interpretation

New measurement modes, a comparison with 2017, and practical care when using DLS results to support decisions.

HajaLux · 27 October 2026

HajaLux editorial illustration of DLS and PdI in ISO 22412:2025.
Editorial illustration; does not depict a specific instrument.

ISO 22412:2025, the third edition published in September 2025, replaces the 2017 edition. Its scope remains average hydrodynamic size and the distribution of mainly submicrometre particles, emulsions and fine bubbles in liquids, covering dilute and concentrated suspensions. It is not restricted to pharmaceutical products.

ISO 2017 vs 2025: comparison

ISO 2017 vs 2025: comparison
Topic ISO 22412:2017 — previous ISO 22412:2025 — current
Status Second edition; replaced. Third edition; current reference.
Measurement angles Optical configurations and angles already considered. MADLS: Clause 9.4.
Imaging DLS No dedicated section in the contents. IDLS and UIDLS: Clause 9.5.
Polarisation and backscatter Cross-correlation already mentioned. PsB PCCS: new A.5.
Concentrated suspensions Already covered; 7.2–7.3. Expanded good-practice guidance.
Quality and uncertainty Already in Clause 10. Clause 10 retained; additional guidance.
ISO/TR 22814 Separate guidance document. Content incorporated into Annexes C and D.
Online measurements Annex C. Expanded Annex B.
Results and reporting Cumulants, distributions and reporting. Framework retained; new modes included.

Original HajaLux table, reconstructed from the requested DSDP Analytics comparison and checked against the ISO preview foreword and contents. An important correction: system qualification and quality control already had a section in 2017; they were not introduced in 2025. The cited annexes are informative. ISO 2025 — preview · ISO 2017 — preview.

Interpreting PdI

Global PdI and peak-specific metrics

Cumulants PdI describes the overall broadening observed in correlation analysis. Its usual expression is PdI = μ₂/Γ̄² (Γ̄: mean decay rate; μ₂: second cumulant.), dimensionless. By contrast, (standard deviation/mean)² calculated for a peak describes that population alone and depends on the chosen distribution. Label it separately; do not present it as global cumulants PdI.

What cannot be attributed to the new edition

The expression “special PdI” appears in manufacturer guidance about customised fitting. The previews consulted do not establish that ISO 2025 created a regulatory category with this name, a “clinical PdI”, or a new general acceptance limit. They also do not support a redefinition of the basic formula.

MADLS and imaging DLS

Including new modes in the standard makes it necessary to identify the configuration and algorithm used. Multiple angles may add information; they do not guarantee universally more accurate PdI. Imaging DLS analyses optical fluctuations and must not automatically be equated with counting or directly measuring the geometry of each particle.

Concentration, turbidity and interpretation

A change after dilution may involve particle interactions, multiple scattering or changes to the formulation itself. Assessment should distinguish these effects. Specific optical features do not automatically make every concentrated sample suitable.

Sources: Cumulants and statistical measures · PdI from an individual peak in DLS.

PdI — Interpreting PdI
PdI Practical reading Interpretation limit
< 0.07 Historical reference for monodisperse spherical standards. Does not certify the quality of every product.
0.08–0.20 Range frequently used in formulation studies. Not a universal ISO criterion or proof of LNP stability.
> 0.70 Warning of a very broad distribution for conventional DLS. Investigate data and method suitability; not automatic regulatory failure.

These ranges are interpretation references, not a continuous acceptance scale. Define criteria for the validated application and procedure. Low PdI alone does not establish storage stability.

Sources: ISO 22412:2017 — preview, 3.3 · DLS — common terms defined.

Terminology

Term / abbreviation Meaning Why it matters
DLS Dynamic light scattering. Links temporal scattering fluctuations to sample dynamics.
Hydrodynamic diameter Size of an equivalent sphere with the same diffusion. Includes the particle response in its medium.
Z-average Hydrodynamic mean obtained by cumulants. Not a number-weighted mean.
PdI / PI Polydispersity index; dimensionless. Describes width in the context of the algorithm.
Cumulants Mathematical expansion of the correlation function. Produces global mean and PdI.
MADLS Multi-angle dynamic light scattering. Report angles and processing.
IDLS / UIDLS Image-based / ultrafast image-based DLS. Does not automatically imply tracking individual particles.
PsB PCCS Polarisation-separated backscatter photon cross-correlation spectroscopy. A specific optical configuration.
Backscatter Light scattered towards the source side. Detection geometry, not a quality index.
Multiple scattering Light scattered more than once before detection. Can change interpretation.
Intensity / volume / number Different distribution weighting bases. Do not compare them as equivalents.
RM / CRM Reference material / certified reference material. Support controls and traceability.
Qualification / validation Checking the instrument / demonstrating procedure suitability. Distinct assessments.
QC Quality control. Tracks result consistency.
CMC / LNP Chemistry, manufacturing and controls / lipid nanoparticle. Documentation context / formulation type.

ASTM E3247: a complementary reference

The ASTM E3247-26 standard related to ISO addresses DLS for nanoparticles in aqueous media, in batch mode. Remember that citing ISO does not automatically demonstrate compliance with ASTM, or vice versa. We will discuss this ASTM standard on another occasion.

Need to review your DLS method?

Discuss the sample, analytical objective and documents to be updated with HajaLux.

Explore our servicesExplore NanoCat Online

References and consultation limits

The links below allow source checking. Tables are original editorial summaries, not reproductions of the standards. Consultation was limited to public material; full clauses, tolerances and conformity conditions must be checked in an authorised complete edition.

  1. ISO 22412:2025 — official catalogue
  2. ISO 22412:2025 — public preview
  3. ISO 22412:2017 — public preview
  4. DSDP Analytics — comparison requested
  5. ASTM E3247-20 — historical edition
  6. ASTM E3247-26 — active edition (27/10/2026)
  7. Malvern Panalytical — statistical measures and cumulants
  8. Malvern Panalytical — PdI from an individual peak
  9. Malvern Panalytical — DLS common terms